Part Overview

The EML30US12-E is an enclosed AC/DC converter manufactured by XP Power, designed for medical applications requiring single-output 12V power delivery at 30W capacity. This unit features universal input voltage range (85–264 VAC) and is rated for 2.5A maximum output current. The product maintains active status with full compliance to medical device standards (IEC 60601-1) and carries CB, TUV, and UL approvals.

Substitute parts become necessary when the primary part experiences extended lead times, inventory constraints, or when system design flexibility permits alternative qualified sources while maintaining electrical and mechanical compatibility.

Substiute Parts

EML30US12-E
XP PowerIn Stock: 1074EML30US12-E Datasheet
EML30US12-E
Current Part
IRM-30-12
MEAN WELL USA Inc.In Stock: 1000502IRM-30-12 Datasheet
IRM-30-12
MFR Recommended

Key Parameters

Parameter Value Substitution Relevance
Power Rating 30 W Critical – must match exactly
Output Voltage 12V Critical – must match exactly
Output Current (Max) 2.5A Critical – must match exactly
Input Voltage Range 85–264 VAC Critical – must match exactly
Number of Outputs 1 Critical – must match exactly
Package Type 4-DIP Module Critical – must match for PCB mounting
Mounting Type Through Hole Critical – must match for assembly compatibility
Operating Temperature Range -20°C to 70°C (with derating) Important – application-dependent
Line Regulation ±0.5% Important – output stability parameter
Load Regulation ±1% Important – output stability parameter
RoHS Status ROHS3 Compliant Important – regulatory compliance

Substitute Part Grouping Explanation

Substitution eligibility for the EML30US12-E is determined by strict equivalence across the following electrical and mechanical parameters:

Electrical Equivalence Requirements:

  • Output voltage: 12V (exact match)
  • Output current capacity: 2.5A (exact match)
  • Power rating: 30W (exact match)
  • Input voltage range: 85–264 VAC (exact match)
  • Single output configuration (exact match)

Mechanical Equivalence Requirements:

  • Package type: 4-DIP Module (exact match)
  • Mounting method: Through Hole (exact match)
  • Physical dimensions: Compatible with PCB footprint constraints

Regulatory and Compliance Requirements:

  • RoHS3 compliance (mandatory for equivalent substitution)
  • Active product status (current manufacturing and support)

The IRM-30-12 manufactured by MEAN WELL USA Inc. satisfies all electrical and mechanical equivalence criteria and maintains active product status with full RoHS3 compliance.

Parameter Comparison

Parameter EML30US12-E (XP Power) IRM-30-12 (MEAN WELL USA Inc.) Match Status
Power Rating 30 W 30 W Equivalent
Output Voltage 12V 12V Equivalent
Output Current (Max) 2.5A 2.5A Equivalent
Input Voltage Range 85–264 VAC 85–264 VAC Equivalent
Number of Outputs 1 1 Equivalent
Package Type 4-DIP Module 4-DIP Module Equivalent
Mounting Type Through Hole Through Hole Equivalent
Operating Temperature Range -20°C to 70°C -30°C to 70°C IRM-30-12 has extended low-temperature range
Efficiency 80% 88% IRM-30-12 has higher efficiency
Line Regulation ±0.5% ±0.5% Equivalent
Load Regulation ±1% ±0.5% IRM-30-12 has tighter tolerance
RoHS Status ROHS3 Compliant ROHS3 Compliant Equivalent
Product Status Active Active Equivalent
Size / Dimension 3.10" L × 1.50" W × 1.10" H (78.7mm × 38.1mm × 27.9mm) 2.74" L × 1.54" W × 0.94" H (69.5mm × 39.0mm × 24.0mm) IRM-30-12 is more compact

Engineering Selection Recommendations

Primary Part: EML30US12-E

The EML30US12-E remains the specified component when medical device certification to IEC 60601-1 is the primary requirement. This unit carries UL approval specifically for medical applications and maintains the exact performance specifications required by the original design.

Substitute Part: IRM-30-12

The IRM-30-12 qualifies as an electrical and mechanical equivalent for applications where:

  1. Medical device certification is not required, or alternative compliance pathways are acceptable
  2. The application operates within the extended temperature range (-30°C to 70°C)
  3. Improved efficiency (88% vs. 80%) and tighter load regulation (±0.5% vs. ±1%) provide system benefits
  4. Compact form factor (69.5mm × 39.0mm × 24.0mm vs. 78.7mm × 38.1mm × 27.9mm) accommodates space constraints
  5. Both units maintain active product status with full RoHS3 compliance and universal input capability

Certification Considerations:

The EML30US12-E carries CB, TUV, and UL approvals with specific reference to IEC 60601-1 medical standards. The IRM-30-12 carries CB, CE, cURus, and TUV approvals. Selection between these units depends on the regulatory framework and application classification of the end system.

Frequently Asked Questions (FAQ)

Q: Can the IRM-30-12 be used as a direct replacement for the EML30US12-E in medical device applications?

A: Electrical and mechanical compatibility is confirmed. However, medical device applications must verify that the IRM-30-12 certifications (CB, CE, cURus, TUV) satisfy the specific regulatory requirements of the target market. The EML30US12-E carries explicit IEC 60601-1 medical device certification; the IRM-30-12 does not list this standard. Regulatory assessment is required before substitution in medical applications.

Q: Are the physical dimensions compatible with existing PCB layouts?

A: Both units use 4-DIP Module packaging with Through Hole mounting. The IRM-30-12 is more compact (69.5mm × 39.0mm × 24.0mm) compared to the EML30US12-E (78.7mm × 38.1mm × 27.9mm). PCB layouts designed for the EML30US12-E will accommodate the IRM-30-12 without footprint conflicts. However, verify clearance around the smaller unit to ensure no design constraints are violated.

Q: Do both units support the same input voltage range?

A: Yes. Both the EML30US12-E and IRM-30-12 accept universal input voltage from 85–264 VAC, enabling operation across global power standards without modification.

Q: What is the difference in output regulation between these two units?

A: The EML30US12-E specifies ±1% load regulation, while the IRM-30-12 specifies ±0.5% load regulation. The IRM-30-12 provides tighter output voltage stability under varying load conditions. Both units maintain ±0.5% line regulation. Selection depends on whether the application requires the improved load regulation performance of the IRM-30-12.

Q: Are both units RoHS3 compliant?

A: Yes. Both the EML30US12-E and IRM-30-12 are ROHS3 compliant, meeting current environmental and hazardous substance restrictions for electronic components.

Q: What is the operating temperature difference between these units?

A: The EML30US12-E operates from -20°C to 70°C with derating, while the IRM-30-12 operates from -30°C to 70°C with derating. The IRM-30-12 extends the low-temperature operating range by 10°C, suitable for applications requiring enhanced cold-environment performance.

Q: Which unit has better efficiency?

A: The IRM-30-12 achieves 88% efficiency compared to the EML30US12-E at 80%. The 8% efficiency improvement reduces thermal dissipation and power loss, benefiting applications with thermal or energy consumption constraints.